Abstract
Temporal variability and reproducibility of Doppler-derived variables obtained during supine symptom-limited exercise was investigated in 26 patients who were in clinically and hemodynamically stable condition with normally functioning nonstented aortic bioprostheses (stentless porcine, n = 13; cryopreserved homografts, n = 13). All patients had normal systolic left ventricular function and underwent 2 similar exercise tests within 12 months (mean time interval 7.2 ± 1.9 months). The coefficient of variation was 8% to 9% for primary Doppler-derived variables (ie, velocities and velocity time integrals) at rest and during exercise. The coefficient of variation for calculated maximal pressure difference was 16% at rest and 15% at peak exercise. Measurement variability assessed from repeated measurements from the same videotaped recording was approximately 2%. High reproducibility was shown for most variables with intraclass correlation coefficients of 0.85 or more. We conclude that Doppler echocardiography can be used in patients with nonstented aortic bioprostheses with the same high reproducibility during exercise as at rest. The results provide clinically useful information regarding temporal variability for Doppler-derived variables. (J Am Soc Echocardiogr 1999;12:484-91.)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society of Echocardiography
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.