Abstract
A new analysis method for the rtOSL of BeO ceramics is presented, using temporal curve fitting of an expected rtOSL signal to measured rtOSL signals. The presented technique does not require heavy signal averaging to determine the OSL bleaching correction associated with the ΔrtOSL method, reducing uncertainties in the post-correction rtOSL. The corrected rtOSL signal was demonstrated to be linear with dose, and dose-rate independent. The presented technique is expected to be applicable for many other dosimeters capable of the rtOSL technique. The presented technique achieved relative uncertainties in the corrected rtOSL between 3.4% and 6.5%. The initial measurements are promising, but uncertainties are required to be further improved upon before the technique can be used clinically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.