Abstract

The aim of this study was to evaluate temperature variation induced by a diode laser in periodontal repair. A diode laser operating with medium power can constitute an alternative device for bacterial reduction at periodontal pockets. This procedure will probably avoid any kind of undesirable thermal damage to the irradiated tissues. The temperature variation induced by a 810-nm diode laser was investigated in an in vitro study, varying the soft tissue thickness, and in an in vivo study for soft periodontal and bone tissues. The laser powers used were 600 mW, 800 mW, 1.0 W, and 1.2 W, and the light was delivered by a 300-microm fiber. The laser parameters and irradiation time used did not induce a temperature variation high enough to cause thermal irreversible damage to the periodontal tissues investigated. This study contributes to the establishment of thermally safe working parameters of a diode medium power laser.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.