Abstract
Many types of ophthalmic instruments produce a concentrated focal irradiance in the lens. Instruments that illuminate large areas of the retina-known as "Maxwellian-view," are but one example, and there are concerns about the potential hazards associated with this optical system. The transfer of the heat generated in the human eye in Maxwellian-view illumination or similar focal-beam situations was simulated using a mathematical model to determine the temperature elevations induced in the human eye. The maximum temperature rise in the lens region was examined to quantitatively assess the potential thermal hazard to the lens. It was shown that Maxwellian-view illumination or similar focal-beam situations can cause thermal injury to the lens under certain conditions, and that this hazard is greater for incident wavelengths of about 320-420 nm than for longer wavelengths. The risk of thermal injury increases as exposure duration increases, and the risk tends to increase as the beam waist diameter or Maxwellian-view angle decreases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.