Abstract

Abstract The ultimate performance of high resolution electron microscopes depends on a variety of factors including the stability of the temperature surrounding the instrument. Variations in temperature can cause drift of the specimen, the microscope electronics, and the mechanical tolerances in components such as microscope lenses and scan coils. For high resolution work or for image reconstruction that requires acquisition of multiple exposures over extended periods of time, it is critical that the temperature variations be strictly controlled. A typical manufacturer's specification for temperature stability of a room housing a transmission electron microscope (TEM) is a tolerance of <0.5°C/hour with fluctuations of <0.05°C/minute. While modern instrumentation rooms are carefully designed to maintain stable environmental conditions (O'Keefe etal, 2004), it seems unlikely that these stringent specifications are routinely met in practice when the microscope is in operation and personnel are present in the room.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.