Abstract

Thermocouples' leads are often too short, necessitating the use of an extension lead. To determine if temperature measures were influenced by extension-lead use or lead temperature changes. Descriptive laboratory study. Laboratory. Experiment 1: 10 IT-21 thermocouples and 5 extension leads. Experiment 2: 5 IT-21 and PT-6 thermocouples. In experiment 1, temperature data were collected on 10 IT-21 thermocouples in a stable water bath with and without extension leads. In experiment 2, temperature data were collected on 5 IT-21 and PT-6 thermocouples in a stable water bath before, during, and after ice-pack application to extension leads. In experiment 1, extension leads did not influence IT-21 validity (P = .45) or reliability (P = .10). In experiment 2, postapplication IT-21 temperatures were greater than preapplication and application measures (P < .05). Extension leads had no influence on temperature measures. Ice application to leads may increase measurement error.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.