Abstract

Single particles trapped in an optical trap may experience temperature elevation, yet direct measurement of temperature and its distribution inside the optical trap of several to hundreds of microns in size remains a big challenge. We introduce a method that can measure the temperature inside a universal optical trap (UOT) using Raman spectroscopy of single trapped particles of high thermal conductivity. We measured temperature and temperature distributions inside the UOT using Raman shifts of single-walled carbon nanotubes (SWCNTs) and micron-sized diamonds (MSDs), which are heated by trapping laser beams directly or indirectly, depending on the location of the particle in the trap. We show that the temperature at the center of the UOT is much lower than the temperature along the hollow beams that form a hollow, cage-shaped UOT. In the range of the trapping laser power of 200-2950 mW, the surface temperature of particles trapped at the center of a UOT changes from 322 K to 830 K, correspondingly. This result gives a heating rate as a high thermal-absorbing particle trapped in the center of the UOT with 18.3 ± 0.4 °C/100 mW. In addition, the temperature gradient outside the UOT was also characterized by trapping SWCNT particles outside the UOT. Results show that when a light-absorbing particle is trapped for the study of material property, phase transitions, surface equilibrium process, chemical reactions, etc., this method can be used to measure temperature distribution and its variations in the trap and its surroundings.