Abstract
In this study a scanned focused ultrasound (SFUS) system was used to heat 66 tumours at various anatomical locations in 52 patients. A total of 160 treatments were given. On average, temperatures were measured in 14 or 15 locations in the scanned volume. The time-averaged temperatures over the 30 min treatment period in the best treatment of each tumour were 44.0 +/- 2.4 degrees C (mean +/- SD) and 39.6 +/- 1.5 degrees C at the location of the highest and lowest sensor, respectively. On average, 39% of the sensors were above 42.5 degrees C. When only the cases that were judged to be good candidates for the hyperthermia device were analysed, 64% of the sensors reached a temperature over 42.5 degrees C with the highest temperature achieved being 45.9 +/- 2.3 degrees C and the lowest 40.7 +/- 1.4 degrees C. Although the system tested has many technical limitations (for example, fixed frequency, beam geometry and power during the scan cycle), the results demonstrate that therapeutic temperatures can be achieved in many tumours. Significantly better temperatures are expected when all of the theoretical potential of scanned focused ultrasound systems has been used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.