Abstract
Helium gas at room temperature is known to diffuse through the epoxies commonly used in various low temperature applications, which can complicate leak detection. The helium flux typically decreases with decreasing temperature. We have measured the flux of helium that passes though thin sections of as-cast clear Stycast 1266, Stycast 2850FT (black) and TRA-BOND 2151 (blue) epoxies as a function of temperature in the range 130K < T < 300K. We analyze the data to create normalized (to constant sample thickness and pressure differential) data for comparison. We report the preliminary temperature-dependent fluxes we have measured, which show significant differences among the epoxies studied.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.