Abstract

The thermal lag of DSC is a heating rate independent constant and influences the measured onset temperature of melting peaks and must be taken into account during the temperature calibration. In the literature, however, non-linear behavior of the measured onset temperature is reported for the case of high scanning rates and the use of small samples. Based on a conventional heat transfer model of a DSC a general relation of the heating rate and sample mass dependence of the onset temperature is derived. The reason of this unusual behavior is that the time for the melting event is less than the equilibration time of the DSC. The approach presented can be used to correct the onset temperatures in the case of non-linear behavior, which may occur in conventional DSC and Fast DSC (using chip sensors).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.