Abstract

The Co–C eutectic fixed point has been popular for thermocouple calibration since 2010. In this study, the ITS-90 temperature of a Co–C eutectic fixed-point cell intended for thermocouple calibration of noble thermocouples was measured using a radiation thermometer and a group of Pt/Pd and PtRh10%-Pt thermocouples. Satisfactory consistency was observed for measurements using both methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.