Abstract

ABSTRACTA novel fiber optic sensor for the simultaneous measurement of refractive index and temperature is reported. The sensor consists of a high-birefringence fiber loop mirror and a section of single mode-coreless-single mode fiber structure. The single mode-coreless-single mode fiber structure served as a refractometer while the high-birefringence fiber loop mirror was used to measure temperature. The multimode interference valley of the single mode-coreless-single mode fiber structure was sensitive to the surrounding refractive index of liquids (96.42 nm/refractive index unit) and had almost no response to temperature fluctuations. The high-birefringence fiber loop mirror was highly sensitive to temperature (1.98 nm/°C) but was insensitive to changes in refractive index. The theoretical and experimental results demonstrated simultaneous measurement of temperature and refractive index. The optimum resolution was 2.07 × 10−4 refractive index units and 0.01°C.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call