Abstract

One of the most commonly employed instruments in the category of handheld condensation particle counters (CPCs) is the TSI 3007, which employs a simplified system for achieving a temperature difference between its saturator and its condenser in view of increasing portability, energy efficiency and autonomy. As a result, particle growth and consequently the detection efficiency of the instrument can be affected by the measurement conditions. In this work we measure the detection efficiency of the TSI 3007 CPC under temperatures and pressures that differ from standard conditions. Our results show that the performance of the CPC has a strong temperature dependence in the range of 5–30 °C, whereas it is not affected by ambient pressure when this varies between 0.7 and 1.0 atm. The temperature dependent detection efficiency of the instrument becomes significant at sizes below 20 nm. Recording the temperature at which this CPC is operated is therefore strongly advised, especially when required to determine the number concentration of sub-20 nm aerosol particles.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.