Abstract

The spatial resolution and the indexing quality obtained with an automated orientation and phase mapping tool are analyzed for different Transmission Electron Microscope (TEM) illumination settings. The electron probe size and convergence angle are studied for two TEM configuration modes referred as microprobe and nanoprobe modes. Using a 10μmC2 aperture in a FEI Tecnai F20 (S)TEM, the nanoprobe mode is used to get a small convergent electron beam while the microprobe mode provides a nearly parallel illumination at the cost of a larger probe size. The nanoprobe configuration enables to increase the spatial resolution (∼1nm vs 3nm) but also affects the fraction of mis-indexed points (15% vs 1%). Indexing errors are attributed to the increase by a factor of three of the convergence angle with respect to the microprobe mode. While intermediate optimum settings may be found and are potentially achievable on electron microscopes providing a ‘free lens’ control or a larger choice of C2 apertures, it is emphasized that the spatial resolution cannot be considered without reference to the indexing quality and, consequently to the convergence angle.