Abstract

An epineural electrode (ENE) was developed to achieve a controllable, progressive muscle contraction and reduce muscle fatigue during functional electrostimulation. In vitro studies of annular (torus-shaped) electrodes have been carried out to select adequate electrode materials and electrode diameters. We investigated the long-term stability of ENEs using different intensities of current (range: 0.8-1.2 mA) in sheep and rat experiments. Depending upon size and material, ENE tolerates anodic and cathodic impulses of up to 1 msec and 8 mA (peak current). ENEs were used during clinical application in four paraplegic patients (16 channel implants) for stimulation of lower extremities, and in nine quadruplegic patients (eight channel implants) for diaphragm pacing, for a total observation time of 37 patient years. The low rate of clinical complications, including no corrosion of the electrodes or significant loss of nerve fibers, seems related to the small dimensions of the electrode (less than 1 mm2).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.