Abstract

Based on in-depth analysis of the relative radiation scaling theorem and acquired scaling data of pixel response nonuniformity correction of CCD (charge-coupled device) in spaceborne visible interferential imaging spectrometer, a pixel response nonuniformity correction method of CCD adapted to visible and infrared interferential imaging spectrometer system was studied out, and it availably resolved the engineering technical problem of nonuniformity correction in detector arrays for interferential imaging spectrometer system. The quantitative impact of CCD nonuniformity on interferogram correction and recovery spectrum accuracy was given simultaneously. Furthermore, an improved method with calibration and nonuniformity correction done after the instrument is successfully assembled was proposed. The method can save time and manpower. It can correct nonuniformity caused by other reasons in spectrometer system besides CCD itself's nonuniformity, can acquire recalibration data when working environment is changed, and can also more effectively improve the nonuniformity calibration accuracy of interferential imaging

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.