Abstract

Detailed analysis of the fetal electrocardiogram (ECG) during labour could provide valuable additional information to assist clinicians in reducing incidents of unnecessary medical intervention without altering current patient handling routines, but poor signal quality and difficulty of processing the ECG without significant distortion have impeded such analysis. The authors present novel optimum curve fitting and digital filtering techniques that make it possible to enhance the quality of the ECG with minimal distortion and to measure, on a beat-by-beat basis, almost any feature (e.g. time constants, amplitudes, areas, maxima and minima) from the fetal ECG waveform accurately. This development removes a major obstacle in the use of fetal ECG and creates a basis for a thorough analysis of the changes in the key fetal ECG features and for evaluating their clinical value in fetal monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.