We investigated the possible origin of the spuriously high results observed with the Abbott TDx Immunoassay in the 1991 Belgian external quality assessment scheme for digoxin. The present work ascribes this discrepancy to a matrix effect induced by the addition of merthiolate as preservative to the control samples. It consequently stresses the importance of avoiding the use of this compound for preparing such samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.