Abstract

In the reconstruction of volume breast images from x-ray projections in breast tomosynthesis, some tomographic systems truncate the image data presented to the radiologist such that a non-negligible amount of tissue may be missing from the breast image. QC tests were conducted to determine if this problem existed in imaging in the TMIST study. Test tools developed for TMIST containing small objects at known heights were used in routine weekly and annual QC testing of tomosynthesis units to assess the degree to which phantom material that was irradiated in imaging was excluded from the reconstructed image. Results from 318 tests on five system types from three manufacturers are reported. The presence and extent of this problem varied among system types. The cause was most frequently related to machine errors in the determination of breast thickness or to deflection of components during breast compression. In particular, the problem occurred when a compression paddle other than the one calibrated for tomosynthesis was used for the tests. This was also verified to have occurred in some clinical imaging. Missing volume can be avoided by intentionally reconstructing additional image slices above and below the presumed locations of the breast support and compression plate. A compression paddle which has been calibrated for tomosynthesis should be used both for clinical imaging and testing. The prevalence of this phenomenon suggests that more frequent testing for volume coverage may be advisable.