Abstract

PurposeSynchrony is a motion management system available on the Radixact linear accelerator that utilizes kilovoltage (kV) radiographs to track target motion and synchronize the delivery of radiation with the motion. Proper management of this imaging dose requires accurate quantification. The purpose of this work was to use Monte Carlo (MC) simulations to quantify organ‐specific patient doses from these images for various patient anatomies.MethodsPoint doses in water were measured per TG‐61 for three beam qualities commonly used on the Radixact. The point doses were used to benchmark a model of the imaging system built using the Monte Carlo N‐Particle (MCNP) transport code. Patient computed tomography (CT) datasets were obtained for 5 patients and 100 planar images were simulated for each patient. Patient dose was calculated using energy deposition mesh tallies.ResultsThe MCNP model was able to accurately reproduce the measured point doses, with a median dose difference of less than 1%. The median dose (D50%) to soft tissue from 100 radiographs among the 5 patient cases ranged from 2.0 to 4.6 mGy. The max dose (D1%) to soft tissue ranged from 6.2 to 31.0 mGy and the max dose to bony structures ranged from 20.2 to 71.7 mGy. These doses can be scaled to estimate total patient dose throughout many fractions.ConclusionsPatient dose is largely dependent on imaging protocol, patient size, and treatment parameters such as fractionation and gantry period. Organ doses from 100 radiographs (an approximate number for one fraction) on the Radixact are slightly less than the doses from Tomo MVCT setup images. Careful selection of clinical protocols and planning parameters can be used to minimize risk from these images.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call