Abstract
This work describes an inexpensive holder for corrosion experiments, developed to avoid crevice corrosion during pitting corrosion tests of corrosion-resistant alloys. It is based on a flooded gasket, incorporated in a Stern-Makrides sample holder. Unlike the Avesta cell designed by Qvarfort, the proposed sample holder does not require a pump. The sample holder attaches to prismatic or cylindrical specimens cut from bars, plates, or any other component with a thickness greater than 10 mm. The sample holder is versatile and with no limitations on chloride concentration, solution composition, or temperature. The testing solution can be deaerated or saturated with any gas if needed. The sample holding method was validated by cyclic potentiodynamic polarization experiments on Types 304, 316 stainless steel, and C-22HS nickel-based alloy. Crevice corrosion was not evident in any of the specimens tested with the proposed method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.