Abstract
The purpose of this study is to obtain the dosimetric parameters of a new Co-60 source used in high dose rate brachytherapy and manufactured by BEBIG (Eckert & Ziegler BEBIG GmbH, Germany). The Monte Carlo method has been used to obtain the dose rate distribution in the updated TG-43U1 formalism of the American Association of Physicists in Medicine. In addition, to aid the quality control process on treatment planning systems (TPS), a two-dimensional rectangular dose rate table, coherent with the TG-43U1 dose calculation formalism, is given. These dosimetric data sets can be used as input data of the TPS calculations and to validate them.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.