Abstract

Stereotactic body radiotherapy (SBRT) has been proven to be beneficial for several disease sites in the (lower) abdomen. However, the quality of the treatment plan, based on a single planning computed tomography (CT), can be compromised due to large inter-fraction motion of the target and organs at risk (OARs) in this anatomical region. The aim of this study was to investigate the feasibility of online adaptive SBRT treatments on a robotic radiosurgery system and to record estimated total treatment times. For two disease sites, locally advanced pancreatic cancer (LAPC) and oligometastatic lymph nodes, four patients with repeat CTs were included in the feasibility study. Quick treatment plan templates were generated based on the planning CT and validated by running them on the plan and fraction CTs. For two cases a dummy run was performed and the individual steps were timed. Dose delivery was the largest contributor to the total treatment time, followed by contour adaptation. Running the quick plan templates resulted in plans similar to unrestricted plans, obeying the OAR constraints. The dummy runs showed that online adaptive treatments were completed in 64 to 83min respectively for oligometastases and LAPC, comparable to other clinically available solutions. This study showed the feasibility of online re-planning for two challenging disease sites within a clinically acceptable time frame on a robotic radiosurgery system, making use of commercially available elements that are not integrated by the vendor.