Abstract

PurposeThe imaging performance and dose of a mobile CT scanner (Brainlab Airo®, Munich, Germany) is evaluated, with particular consideration to assessment of technique protocols for image‐guided brachytherapy.MethodDose measurements were performed using a 100‐mm‐length pencil chamber at the center and periphery of 16‐ and 32‐cm‐diameter CTDI phantoms. Hounsfield unit (HU) accuracy and linearity were assessed using materials of specified electron density (Gammex RMI, Madison, WI), and image uniformity, noise, and noise‐power spectrum (NPS) were evaluated in a 20‐cm‐diameter water phantom as well as an American College of Radiology (ACR) CT accreditation phantom (Model 464, Sun Nuclear, Melbourne, FL). Spatial resolution (modulation transfer function, MTF) was assessed with an edge‐spread phantom and visually assessed with respect to line‐pair patterns in the ACR phantom and in structures of interest in anthropomorphic phantoms. Images were also obtained on a diagnostic CT scanner (Big Bore CT simulator, Philips, Amsterdam, Netherlands) for qualitative and quantitative comparison. The manufacturer’s metal artifact reduction (MAR) algorithm was assessed in an anthropomorphic body phantom containing surgical instrumentation. Performance in application to brachytherapy was assessed with a set of anthropomorphic brachytherapy phantoms — for example, a vaginal cylinder and interstitial ring and tandem.ResultNominal dose for helical and axial modes, respectively, was 56.4 and 78.9 mGy for the head protocol and 17.8 and 24.9 mGy for the body protocol. A high degree of HU accuracy and linearity was observed for both axial and helical scan modes. Image nonuniformity (e.g., cupping artifact) in the transverse (x,y) plane was less than 5 HU, but stitching artifacts (~5 HU) in the longitudinal (z) direction were observed in axial scan mode. Helical and axial modes demonstrated comparable spatial resolution of ~5 lp/cm, with the MTF reduced to 10% at ~0.38 mm−1. Contrast‐to‐noise ratio was suitable to soft‐tissue visualization (e.g., fat and muscle), but windmill artifacts were observed in helical mode in relation to high‐frequency bone and metal. The MAR algorithm provided modest improvement to image quality. Overall, image quality appeared suitable to relevant clinical tasks in intracavitary and interstitial (e.g., gynecological) brachytherapy, including visualization of soft‐tissue structures in proximity to the applicators.ConclusionThe technical assessment highlighted key characteristics of dose and imaging performance pertinent to incorporation of the mobile CT scanner in clinical procedures, helping to inform clinical deployment and technique protocol selection in brachytherapy. For this and other possible applications, the work helps to identify protocols that could reduce radiation dose and/or improve image quality. The work also identified areas for future improvement, including reduction of stitching, windmill, and metal artifacts.

Highlights

  • Advantageous features include improved image quality compared to cone‐beam CT (CBCT) and reduced cost / increased flexibility compared to a conventional diagnostic multidetector CT (MDCT) scanner in a dedicated simulation room

  • Previous studies reported the performance of the mobile CT scanner within the context of its primary indication in image‐guided spine surgery

  • Their work showed spatial resolution up to 4 lp/cm in the head field‐ of‐view (FOV) for the mobile scanner, compared to 7 lp/cm resolution from a Siemens Sensation 64‐slice MDCT scanner for the same FOV.[1]

Read more

Summary

Introduction

Mobile systems for intraoperative 3D imaging have become prevalent over the last decade. One such system is the Brainlab Airo (Brainlab Airo®, Munich, Germany), a mobile CT scanner based on a 32‐row detector with a large bore size (107‐cm inner diameter), small footprint, and slim gantry design. Advantageous features include improved image quality compared to cone‐beam CT (CBCT) and reduced cost / increased flexibility compared to a conventional diagnostic multidetector CT (MDCT) scanner in a dedicated simulation room. As such systems are introduced, rigorous technical assessment can help guide clinical implementation and development of future applications in image‐guided interventions. The Airo was found to exhibit higher radiation dose than the Sensation 64 for comparable technique factors (50% increase for head, 85% increase for body phantom), and ring‐like artifacts were noted as contributors to increased low‐frequency noise in the image NPS.[1]

Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call