Abstract
We evaluated the analytical and clinical performance of six commercial immunoradiometric assay kits for thyrotropin (TSH) in serum in 218 subjects, with and without thyroid dysfunction. Detection limits of the six kits were lower (from 0.07 to 0.25 milli-int. unit/L) than that of conventional TSH RIA (0.7 milli-int. unit/L). Precision was adequate over a wide range of concentrations, although interassay CVs at very low concentrations were good for only two kits (7.3% and 13.1%). Results by all the kits correlated to about the same degree with the TSH RIA (r = 0.92 to 0.98). All showed a positive correlation between both the basal and post-thyroliberin (TRH)-stimulation values for TSH (r = 0.78 to 0.88), and all showed similar euthyroid reference ranges for basal concentrations of TSH. With four of the six kits we could clearly distinguish most of the hyperthyroid patients from healthy euthyroids; however, basal and post-TRH TSH values were not sufficient to discriminate among groups of patients with different grades of thyroid hyperfunction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Clinical Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.