The aim of this study was to evaluate quantitatively the fracture—by tear and delamination—of bovine pericardium tissues which are usually employed for the manufacture of bioprosthetic valves. A large number of samples (77) were tested in root-to-apex and circumferential directions, according to a standardised tear test (ASTM D 1938). Before performing the tear test, some samples were subjected to 1000 cycles of fatigue to a maximum stress of 3MPa. Fracture toughness of tearing and delamination were computed by following a simple fracture model. The study showed significantly lower values of delamination toughness compared with tear delamination. Moreover, tear forces were different in each test direction, revealing a clear orthotropic behaviour. All these results, as well as the testing procedure, could be of value for future research in the physiological function of pericardium tissues and clinical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.