Abstract

Abstract Taurine/EPR rods (3 × 10 mm) have been prepared by a simple technique in the laboratory where taurine powder was mixed with a molten mixture of paraffin wax and an ethylene vinyl acetate (EVA) copolymer. The binding mixture EVA/Paraffin does not present interference or noise in the EPR signal before or after irradiation. The rods show good mechanical properties for safe and multi-use handling. An EPR investigation of radiation induced radicals in taurine rods revealed that there are two types of radicals produced after exposure to gamma radiation (60Co). EPR spectra were recorded and analyzed - also the microwave power saturation and modulation amplitude were studied and optimized. Response of taurine to different radiation doses (1.5-100 kGy) was studied and found to follow a linear relationship up to 100 kGy. Radiation induced radicals in taurine persists and showed a noticeable stability over 94 days following irradiation. Uncertainities associated with the evaluation of radiation doses using taurine dosimeters were discussed and tabulated. It was found that taurine possesses good dosimetric properties using EPR spectroscopy in high doses in addition to its simple spectrum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.