Abstract

PurposeThe present study aims to evaluate TTFs of deep-learning-based image reconstruction (DLIR) and iterative reconstruction (IR) in computed tomography (CT) using a conventional task with a rod object with a diameter of 30 mm and a newly-proposed task with a wire of 1 mm in diameter, simulating large and small enhanced vessels, respectively. MethodsThe rod or wire phantom made of a material equivalent to diluted iodine that exhibits about 270 Hounsfield unit (HU) was placed inside a 30-cm water phantom. In-plane and z-directional TTFs were measured for the rod using the circular edge (CE) and plane edge (PE) methods, respectively. By using the wire (iodine wire: IW), in-plane and z-directional TTFs were measured using Fourier transform (IW method). TTFs of filtered back projection (FBP), IR, and DLIR of a 256-row CT system and FBP and IR of a 64-row CT system were evaluated with CT dose indices of 10 and 5 mGy. ResultsFor DLIR and IR, TTFs measured using the IW method were notably lower than those using the CE (or PE) method; moreover, they were also lower than those of corresponding FBP, indicating that the small enhanced vessels with a diameter of about 1 mm would be blurred with both DLIR and IR. ConclusionsThe proposed IW method has turned out to be effective to evaluate TTFs for small enhanced vessels, which have not been properly evaluated by the CE or PE method conventionally recommended.