Abstract

MRI-guided prostate needle biopsy requires compensation for organ motion between target planning and needle placement. We propose slice-to-volume registration algorithms for tracking the prostate motion. Three orthogonal intra-operative slices are acquired in the approximate center of the prostate and registered with a high-resolution target planning volume. Both rigid and deformable scenarios were implemented. MRI-guided robotic prostate biopsy cases were analyzed retrospectively. Average registration errors were 2.55mm for the rigid algorithm and 2.05mm for the deformable algorithm. Slice-based tracking appears to be promising. Deformable registration does not seem warranted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.