Abstract

At the IRIS facility a rather precise method of the target-ion source unit ionization efficiency measurement has been developed. The method exploits an off-line mass-separator for the implantation of the ion beams of selected stable isotopes into a tantalum foil placed inside a Faraday cup in the focal plane of the mass-separator. After the implantation of the required amount of the investigated species, tantalum foil has been inserted into the volume of the target-ion source unit prepared for the on-line utilization at the IRIS on-line separator. The first tests have ensured the ionization efficiency values (90±10)% for Rb and (85±10)% for Cs in the empty combined target-ion source unit, which was used as a reference one. For the target-ion source unit with UC target material inside the measured value of the ionization efficiency of Rb was (52±20)%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.