Abstract

A tandem time-of-flight mass spectrometer for the study of photodissociation of biopolymer ions generated by matrix-assisted laser desorption ionization was designed and constructed. A reflectron with linear and quadratic (LPQ) potential components was used. Characteristics of the LPQ reflectron and its utility as the second stage analyzer of the tandem mass spectrometer were investigated. Performance of the instrument was tested by observing photodissociation of [M + H] + from angiotensin II, a prototype polypeptide. Quality of the photodissociation tandem mass spectrum was almost comparable to that of the post-source decay spectrum. Monoisotopic selection of the parent ion was possible, which was achieved through the ion beam–laser beam synchronization. General theoretical considerations needed for a successful photodissociation of large biopolymer ions are also presented.