Abstract

In this work, high-flow online reversed-phase extraction was coupled with normal phase on silica columns with aqueous-organic mobile phase liquid chromatography/tandem mass spectrometry (LC/MS/MS) to quantify drug candidates in biological fluids. The orthogonal separation effect obtained from this configuration considerably reduced matrix effects and increased sensitivity for highly polar compounds as detected by selected reaction monitoring. This approach also significantly improved the robustness and limit of detection of the assays. An evaluation of this system was performed using a mixture of albuterol and bamethan in rat plasma. Assay validation demonstrated acceptable accuracy (< 8% difference) and precision (< 6% CV) for these model compounds. The system has been used for the quantitation of polar ionic compounds in biological fluids in support of drug discovery programs. This assay was used to analyze samples for a BMS proprietary compound (A) in a rat pharmacokinetic study and is shown as an example to demonstrate the precision, accuracy, and sufficient sensitivity of this system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.