Abstract

To develop short water suppression sequences for 7 T magnetic resonance spectroscopic imaging, with mitigation of subject-specific transmit RF field ( B1+) inhomogeneity. Patient-tailored spiral in-out spectral-spatial saturation pulses were designed for a three-pulse WET water suppression sequence. The pulses' identical spatial subpulses were designed using patient-specific B1+ maps and a spiral in-out excitation k-space trajectory. The subpulse train was weighted by a spectral envelope that was root-flipped to minimize peak RF demand. The pulses were validated in in vivo experiments that acquired high resolution magnetic resonance spectroscopic imaging data, using a crusher coil for fast lipid suppression. Residual water signals and MR spectra were compared between the proposed tailored sequence and a conventional WET sequence. Replacing conventional spectrally-selective pulses with tailored spiral in-out spectral-spatial pulses reduced mean water residual from 5.88 to 2.52% (57% improvement). Pulse design time was less then 0.4 s. The pulses' specific absorption rate were compatible with magnetic resonance spectroscopic imaging TRs under 300 ms, which enabled spectra of fine in plane spatial resolution (5 mm) with good quality to be measured in 7.5 min. Tailored spiral in-out spectral-spatial water suppression enables efficient high resolution magnetic resonance spectroscopic imaging in the brain. Magn Reson Med 79:31-40, 2018. © 2017 International Society for Magnetic Resonance in Medicine.