Abstract

The simultaneous acquisition of spatial harmonics (SMASH) imaging technique uses spatial information from an array of RF coils to substitute for omitted encoding gradient steps and thereby to accelerate MR image acquisition. Since SMASH image reconstructions rely on the accurate generation of sinusoidally varying composite sensitivity functions to emulate the spatial modulations produced by gradients, the technique was originally believed to be limited to certain image planes or coil array configurations which were particularly suited to the generation of spatial harmonics. Several key improvements to the SMASH reconstruction procedure are described, taking advantage of various degrees of freedom in the spatial harmonic fit. The use of tailored fitting procedures, in combination with a numerical conditioning approach based on new observations about noise propagation in the fit, are shown to allow high-quality SMASH image reconstructions in oblique and double-oblique image planes, both in phantoms and in high-resolution cardiac MR images. Magn Reson Med 44:243–251, 2000. © 2000 Wiley-Liss, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.