Abstract

At the beginning of the recording, atrial tachycardia (AT) can be observed with 1:1 conduction at ventricular rate 140 bpm (upper track). At the spontaneous termination of the tachycardia, atrial-ventricular pacing at 45 bpm (lower rate) and after two cycles atrial-ventricular pacing at 100 beats/min (upper track) can be observed. The rate drop response was activated: drop size 25 ppm, drop rate 50 ppm, detection beat 2, detection window 1 min, intervention rate 100 ppm, intervention duration 2 min (Fig. 1). Fig. 1 Holter recording of rate drop response induced by paroxysmal atrial tachycardia The termination of atrial tachycardia is followed by 45 bpm pacing, which is lower than the drop rate (50 bpm) while the rate drop is higher than the drop size (25 bpm). These two parameters can activate rate drop response separately and in combination, resulting in 100 bpm pacing. The upper track was reduced to 100 bpm and AT ablation was suggested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.