Abstract
To investigate the accuracy of low signal-to-noise ratio (SNR) T(2) and T(2)* measurements using array coils and optimal B(1) image reconstruction (OBR) compared to the standard root sum of squares (RSS) reconstruction. Calibrated gels were used for the in vitro study of T(2). T(2) and T(2)* measurements were obtained from a volunteer's knee and liver, respectively. T(2) and T(2)* measurements were performed using multiecho spin echo and multiecho gradient echo sequences, respectively. SNR was deliberately kept low. The same raw data were used for both reconstructions. For the in vivo studies the effect of signal averaging was also investigated. The optimal reconstructions demonstrated a lower mean background noise level than RSS. In vitro, the T(2) measurements made with OBR images agreed better with a reference high SNR measurement than measurements made from RSS images; the RSS image results overestimated the T(2.) In vivo, increasing the signal averages decreased the difference between the measurements obtained using the OBR and RSS methods, with RSS resulting in longer relaxation times. This work demonstrates improvements to the accuracy of T(2) and T(2)* measurements obtained when OBR is used compared to RSS, particularly in the case of low SNR.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.