Abstract

To compare T2 relaxation time measurements between MR pulse sequences at 3 Tesla in agar phantoms and in vivo patellar, femoral, and tibial articular cartilage. T2 relaxation times were quantified in phantoms and knee articular cartilage of eight healthy individuals using a single echo spin echo (SE) as a reference standard and five other pulse sequences: multi-echo SE (MESE), fast SE (2D-FSE), magnetization-prepared spoiled gradient echo (3D-MAPSS), three-dimensional (3D) 3D-FSE with variable refocusing flip angle schedules (3D vfl-FSE), and quantitative double echo steady state (qDESS). Cartilage was manually segmented and average regional T2 relaxation times were obtained for each sequence. A regression analysis was carried out between each sequence and the reference standard, and root-mean-square error (RMSE) was calculated. Phantom measurements from all sequences demonstrated strong fits (R(2) > 0.8; P < 0.05). For in vivo cartilage measurements, R(2) values, slope, and RMSE were: MESE: 0.25/0.42/5.0 ms, 2D-FSE: 0.64/1.31/9.3 ms, 3D-MAPSS: 0.51/0.66/3.8 ms, 3D vfl-FSE: 0.30/0.414.2 ms, qDESS: 0.60/0.90/4.6 ms. 2D-FSE, qDESS, and 3D-MAPSS demonstrated the best fits with SE measurements as well as the greatest dynamic ranges. The 3D-MAPSS, 3D vfl-FSE, and qDESS demonstrated the closest average measurements to SE. Discrepancies in T2 relaxation time quantitation between sequences suggest that care should be taken when comparing results between studies.