Abstract

Proton density and transverse relaxation (T2)-weighted fast spin echo images are frequently acquired. T2 quantification is commonly performed by applying an exponential fit to these two images, despite recent evidence that an exponential fit is insufficient to correctly quantify T2 in the presence of imperfect RF refocusing due to standard 2D slice selection or use of reduced refocusing angles. Here we examine the feasibility of accurate two echo fitting using standard proton density and T2-weighted images by utilizing Bloch equation simulations and prior knowledge of refocusing angles. This method is demonstrated in simulation, phantom, and human brain experiments, in comparison to the exponential approach, and to a 32 echo multiple-echo spin echo approach. Comparison to single spin echo is also performed in phantom experiments. The two echo method, which compensates for indirect and stimulated echoes, enables accurate quantitative T2 over a wide range of flip angle and T2 values using standard MRI methods, provided there is adequate SNR and flip angle knowledge.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.