ObjectivesTo intraindividually compare image quality and diagnostic performance of multiparametric MRI (mpMRI) at 3T for the detection of prostate cancer (PCa) using a pelvic phased-array coil (PAC) and a combined endorectal and pelvic phased-array coil (ERC-PAC). MethodsForty-five patients were prospectively included and received mpMRI of the prostate using a PAC and an ERC-PAC during one imaging session. Two radiologists evaluated image quality and the most suspicious lesion according to the PI-RADS scoring system. Results of MRI-TRUS-fusion biopsy of the prostate served as reference standard. Patient comfort and acceptance were assessed using a standardized questionnaire. ResultsOverall image quality for T2WI was rated significantly better with an ERC-PAC compared to a PAC (p=0.0038). The weighted kappa for PI-RADS scores for T2WI and DWI with a PAC and an ERC-PAC was 0.70 and 0.73, respectively. For a PI-RADS sum score including T2WI and DWI the area under the curve with a PAC and an ERC-PAC were 0.95–0.99 and 0.93–0.97, respectively (p=0.1395). ConclusionFor T2WI and DWI performed at 3T index PCa lesion identification and evaluation did not differ significantly with both coil setups. Patients preferred MRI without an ERC. Therefore, the use of an ERC may be omitted in a prostate cancer detection setting.