Abstract

T-wave morphology parameters based on principal component analysis (PCA) are candidate to better understand the relation between QT prolongation and torsades de pointes. We aimed to assess the repeatability and to determine the influence of T-end position on PCA parameters. Digital ECGs recorded from 30 subjects were used to assess short term (5 minutes), circadian and long-term (28 days) repeatability of PCA parameters. The T-end cursor position was moved backward and forward (+/- 8 ms) from its optimal position. We calculated QRS-T angle, PCA ratio, and T-wave residuum (TWR). At long-term evaluation, coefficients of variation were 11.3 +/- 9.9%, 11.7 +/- 7.1%, and 23.0 +/- 22.0% for the QRS-T angle, PCA ratio, TWR, respectively. After moving the T-end cursor, repeatability was 0.42 +/- 0.2%, 1.00 +/- 1.04%, 4.0 +/- 4.2% for the same PCA parameters. T-wave morphology parameters based on PCA are reproducible with the exception of TWR and QRS-T angle. In addition, PCA is robust, showing only little dependence on T-end cursor position. These data should be taken into account for safety pharmacology trials.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.