Abstract
A systematic study is performed about several factors involved in the measurements when a absorbent blank is used, as molar absorption coefficient of the formed complex, absorbance of the blank at the measured wavelength, and kind of spectrophotometer used (conventional double-beam or diode-array). It is proved that the usual methodology, measuring against reagent blank, produces less accurate measurements than those obtained against solvent blank. On the other hand, use of the H-Point standard additions method provides always equal or better results than the best ones obtained at single wavelengths by using measurements against solvent bank, at the maximum of the formed complex. To test the methodology, calcium determination with Arsenazo III, at basic pH, is studied at different peaks.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.