Abstract

Photoluminescence (PL) quantum efficiency (QE) is one of the important parameters to characterize photoluminescent materials especially for lighting and display applications. Under a few assumptions, IEC standard (IEC62607-3-1) provides two methods of high simplicity and moderate accuracy with an integrating sphere for absolute measurement of QE named ‘collimated incident light method (CILM)’ and ‘diffuse incident light method (DILM)’, respectively (International standard IEC 62607-3-1 2014 Nanomanufactureing-Key Control Characteristics Part3-1: Luminescent Nanomaterials-Quantum Efficiency 1st edn (Geneva: International Electrotechnical Commission); de Mello et al 1997 Adv. Mater. 9 230–2). However, the simplification unavoidably causes systematic errors, which could be significant in certain cases of sphere reflectance or material properties. In this paper, we numerically investigate the systematic errors that can occur in the standardised methods when employing a Monte Carlo ray-tracing simulation and also present the correction method based on additional measurement to estimate the diffusivity of the sample reflectance at excitation wavelength and absorption of the sample at PL emission wavelength.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call