Abstract

This paper presents an approach to systematic error correction for improving the accuracy of the electrosurgical unit (ESU), through a method that enables safer operation for both surgeons and patients. The equations for systematic errors encountered during the calibration of the eight ESUs were obtained using regression analysis. The ESU calibration was performed using a power analyzer and the reference power value was selected on the ESU panel; the power value indicated by the ESU was noted from the power analyzer and recorded. The fitted regression models were suitable for predicting systematic errors associated with ESU output power values, as a reduction of about 84 % was noted in cutting power (CuP) and about 74 % for coagulation power (CoP). After correction, systematic error values remained below 1.9 W and 0.4 W in cutting and coagulation power, respectively. The maximum error values obtained were less than those specified for the IEC 60601-2-2 standard for all evaluated ESUs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.