Abstract

We compare multiple temporal pulse characterization techniques in three different pulse duration regimes from 15 fs to sub-5 fs, as there are no available standards yet for measuring such ultrashort pulses. To accomplish this, a versatile post-compression platform was developed, where the 100 fs near infrared pulses were post-compressed to the sub-two-cycle regime in a hybrid, three-stage configuration. After each stage, the duration of the compressed pulse was measured with the d-scan, TIPTOE and SRSI techniques and the retrieved temporal intensity profiles, spectrum and spectral phases were compared. Spectral homogeneity was also measured with an imaging spectrometer to understand the input coupling conditions of the temporal measurements. Our findings suggest that the different devices give similar results in terms of temporal intensity profile, however they are extremely sensitive to alignment and to beam quality, especially in the case of the shortest pulses. We address specific steps of measurement procedures, which paves the way towards the standardization of pulse characterization in the near future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.