Abstract

The objective was to characterize both systematic and random errors in Positron Emission Tomography (PET)-based 90 Y three-dimensional (3D) dose quantification. A modified NEMA-IEC phantom was used to emulate 90 Y-microsphere PET imaging conditions: sphere activity concentrations of 1.6and 4.8MBq/cc, sphere-to-background ratios of 4 and 13, and sphere diameters of 13, 17, and 37mm. PET data were acquired using a GE D690 PET/CT scanner for 300min on days 0-11. The data were downsampled to 60-5min for multiple realizations to evaluate the count starvation effect. The image reconstruction algorithm was 3D-OSEM with PSF+TOF modeling; the parameters were optimized for dose-volume histogram (DVH), as a 90 Y 3D dose quantification. 90 Y-PET images were converted to dose maps using the local deposition method, then the sphere DVHs were calculated. The ground truth for the DVH was calculated using convolution method. Dose linearity was evaluated in decaying 90 Y activity (reduced count rate and total count) and decreasing acquisition durations (reduced total count only). Finally, the impacts of the low 32-ppm positron yield on PET-based 3D 90 Y-dose quantification were evaluated; the bias and variability of resulting DVHs were characterized. We observed nonlinear errors that depended on the 90 Y activity (count rate) and not on the total true prompt counts. These nonlinear errors in mean dose underestimated the measured mean dose by>20% for a measured dose range of 40-230Gy; although the shapes of the DVH were not altered. Compensation based on empirical models reduced the nonlinearity errors to be within 5% for measured dose range of 40-230Gy. In contrast, the errors due to nonuniformity introduced by image noise dominated the systematic errors in the DVH and stretched the DVH on both tails. For the 37-mm sphere, the magnitude of errors in D80 increased from -25% to -36% when acquisition duration was decreased from 300 to 10min. The effect of image noise on DVH was more extensive in smaller spheres; for the 17-mm sphere, the magnitude of errors in D80 increased from -29% to -45% acquisition duration was decreased from 300 to 10min. For the 37-mm sphere, the errors in D20 increased from +3.5% to only +10.5% when the acquisition duration was decreased from 300 to 10min; in the 17-mm sphere, the errors in D20 were 6.5% for both 300- and 10-min sphere images. Count-starved 90 Y-PET data introduce both systematic and random errors. The systematic error increases the apparent nonuniformity of the DVH, while the random error increases the uncertainty in the DVH. The systematic errors were larger than the random errors. Lower count rate of 90 Y-PET also introduces systematic bias, which is scanner specific. The errors of bias-compensated mean tumor dose were <10% when 90 Y-PET scan time was >15min/bed for tumors >37mm. Dmedian and Dmean were the most stable dose metrics. An acquisition duration of 30min is recommended to keep the random errors<10% for a typical tumor with sphere equivalent diameter >17mm and average tumor dose >40Gy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.