Abstract

In 2021, the Institute of Electrical Engineering, Chinese Academy of Sciences successfully reached 9.4 T in a whole-body magnetic resonance imaging (MRI) superconducting magnet with an inner diameter of 800 mm. In this study, a systematic analysis of both the real quench protection performance and a simulation are reported. The four successful quench protections during the entire energization process proved the feasibility of the ‘in-out-in’ quench protection protocol for a 9.4 T-800 mm superconducting magnet. The quench trigger sequence was shown to be adjusted by changing the heater thickness, which demonstrates the flexibility of the ‘in-out-in’ quench protection protocol to fit a different MRI magnet design. The high accuracy of the quench protection simulation method and code was confirmed through comparison of the simulation results with the real performance. Moreover, the limitations of the current quench protection and reference value to other MRI magnets were discussed. It is believed that this study will be useful to other research groups and promote the development of an extremely high-field whole-body MRI system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.