Abstract

The simultaneous analysis of diuretic agents by reversed-phase liquid chromatography with a diode-array detector (DAD) was performed by using a gradient elution with acetonitrile and phosphate buffer on a Hypersil-ODS column. For the spiked urine the extraction recovery of solid-phase extraction (SPE) using Sep-Pak C18 cartridge was compared with that of liquid-liquid extraction (LLE) with diethyl ether at various pH. The standard calibration curves were linear from 0.20-20.0 micrograms/mL for all diuretic agents except amiloride, 1.0-20.0 micrograms/mL, and the detection limit was about 0.2 micrograms/mL for 3 mL of urine, except that of amiloride, which was 1.0 micrograms/mL. The confirmation analysis was performed by gas chromatography/mass spectrometry (GC/MS) following methylation. The characteristic mass fragment ions obtained by electron-impact (EI) ionization (70 eV) provided identification of each diuretic agent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.