Abstract

A simple three-channel system for simultaneous determination of As, Se, and Sb by hydride generation and ICP emission spectrometry is described. The device is based on the use of simple, low-resolution prism monochromators for spectral line isolation, together with interference filters for general background reduction. Detection limits are found to be 1.0 μg/L, 1.3 μg/L, and 2.4 μg/L, respectively, for As, Se, and Sb. Interference studies show no significant problems for a large number of elements. Use of the method for the analysis of Standard Reference Water Samples showed generally good agreement with round-robin data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.