Abstract

A system which allows sensitive and fast automated analysis of weakly labeled fluorescent specimens is described. It is tested in the analysis of c- fos expression stimulated by fetal calf serum and calibrated by quantitation of defined solutions injected into cells with the automated microinjection system. Low light level imaging technology combined with quantitative image processing methods and computer control of the hardware allows fully automated analysis of fluorescent molecules in single living or fixed cells. Reliable methods for subtraction of fluorescent background and automated identification of objects of interest in double-stained cells are described. The accuracy of quantitation is considerably improved by normalizing the fluorescence intensities of respective fluorophores in the same object by the method of ratio imaging. The error rate in determining the relative protein content in single cells is less than 15%. The method is applied to microinjection studies with a monoclonal antibody against casein kinase II subunit β. Microinjection of this antibody into synchronized cells specifically inhibits c- fos expression stimulated by fetal calf serum. In combination with the computer-automated capillary microinjection system, the technique will become a useful tool in experiments requiring quantitative single cell analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.