Abstract
Despite the LCD monitor manufacturing costs per square inch dropping below $2 this year, full-size, 14″×17″ digital chest x-ray detector systems are still being sold commercially for tens of thousands of dollars, for two main reasons. First, the extra required p+ doped silicon layer cannot be deposited in the same fabrication facility as a standard thin-film transistor panel for LCD monitors. Secondly, low volume of sales of digital radiography panels result in large margins and high sales prices for detector manufacturers. In this report, we show the system design of a truly low-cost, digital x-ray detector. The primary cost-cutting technique is the replacement of conventional, vertical PIN photodetectors with lateral metal-semiconductor-metal photodetectors, which can be fabricated using existing TFT processes. This paper covers the design of the pixel architecture, pixel array, and readout electronics.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.